Industry

Semiconductor devices

C-V, I-V and low-current measurements on power and optoelectronic semiconductors, from femtoamperes to kilovolts.

TH521 power device analyzer on a lab bench

Power and optoelectronic semiconductors need measurements at both extremes: capacitances of about 1 pF, currents in the femtoampere range and voltages up to several kilovolts. The tasks below cover C-V characterisation of power devices, I-V measurements on photodiodes and MOSFETs, and junction capacitance of photodiodes.

Typical test tasks

Test taskDeviceTypical test conditionsRecommended set-up
C-V characteristics of high-voltage power devicesIGBT modules for wind power, traction and PV invertersDrain-source voltage from 1300 V up to 3000 V; integration into automated linesTH513 C-V analyzer with TH513-1 test fixture and safety interlock
Parasitic capacitance of MOSFETsDiscrete MOSFETs, multi-chip devices and modulesCiss, Coss, Crss, Rg; VGS about 20 V, VDS 0–1000 VTH510 series semiconductor C-V analyzer (TH512 for VDS up to 1500 V)
Dark current of avalanche photodiodes (APD)APDs for optical communication, laser ranging and medical imagingDark current around 0.9 nA; measurement in a shielded set-upTH199X source/measure unit with shielded test box and triaxial cables
MOSFET matchingMOSFET pairs for high-frequency power suppliesThreshold voltage and transconductance curvesTH199X source/measure unit with TH26087B adapter box, TH26087E shielded test box and TH26087G triaxial cables
PN junction capacitance of photodiodesPhotodiode chips and devicesAbout 1 pF at 1–2 MHz, bias 10–40 V; single- or double-sided chip contactingTH2838 precision LCR meter (TH2838H for DC bias up to ±40 V)
Characterisation of power devicesPower MOSFETs and IGBTsIV parameters, capacitance, gate charge and power lossTH521 power device analyzer

What to consider

  • Safety at high voltage. C-V tests at kilovolt levels need fixtures with an interlock, and an analyzer that withstands a device breakdown during the test.
  • Shielding for low currents. Femtoampere and picoampere measurements need shielded test boxes and triaxial cables. Otherwise noise dominates the result.
  • Automation. For production test, check the handler interface and the remote-control protocol (for example SCPI) before you plan the line.

All applications