Digital Multimeters & Data Acquisition
TH1953 / TH1963 / TH1963A Series
Bench Digital Multimeters, 5½ and 6½ digits
A compact bench multimeter family covering three precision classes on one platform, from a general-purpose 5½-digit instrument up to a 6½-digit, 1,199,999-count multimeter with dual front-and-rear input terminals. All three models share the same 4.3" colour display, ten measurement functions and remote-control interfaces, so a production line or lab can standardise on one user interface across every bench.
Technical question about this series? Ask our team.
Key features
- 6½-digit, 1,199,999-count resolution (TH1963 / TH1963A); 5½-digit, 119,999-count on TH1953
- Measurement speed up to 1,000 readings/s
- 4.3" colour LCD with direct reading, histogram, bar graph and trend chart display
- Front and rear input terminals on TH1963 for flexible rack wiring
- AC voltage and current measurement down to 3 Hz
- Built-in capacitance measurement, up to 5 V diode test voltage
- Internal data logging for up to 10,000 readings
- Math functions: Min/Max/Average/Standard deviation, dB, dBm, null/relative
- Pass/fail limit testing with HI/LO/IN comparator and audible beep
- RS232, USB Host, USB Device and LAN standard; GPIB or HANDLER optional
3 models, one platform
TH1953
Entry-level 5½-digit bench DMM
- Resolution
- 119,999 counts
- DC voltage accuracy
- 0.010 + 0.004%
- Input terminals
- Front only
TH1963
6½-digit, front & rear input terminals
- Resolution
- 1,199,999 counts
- DC voltage accuracy
- 0.0035 + 0.0005%
- Input terminals
- Front & rear
TH1963A
6½-digit, front-panel input only
- Resolution
- 1,199,999 counts
- DC voltage accuracy
- 0.0075 + 0.0005%
- Input terminals
- Front only
TH1963 carries a full second set of Sense Ω4W / Input HI-LO and current terminals on the rear panel. Wire the rear terminals once for a rack or fixture and switch to the front terminals for bench use without re-cabling. TH1963A and TH1953 make all connections at the front panel. Accuracy is stated as ±(% reading + % range), TCAL = 25 °C.
Applications
Production and maintenance
Production-line test benches and repair/maintenance workbenches needing fast, repeatable DC/AC voltage, current and resistance readings.
Teaching and laboratory
Teaching laboratories and general R&D bench measurement, where a clear colour display and simple front-panel operation matter.
Automated test systems
ATE integration via RS232, USB, LAN, and optional GPIB or HANDLER for limit-test sorting.
Technical specifications
| Parameter | TH1953 | TH1963 | TH1963A |
|---|---|---|---|
| Display | 4.3" LCD colour display, English menu | ||
| Display digits | 119,999 counts | 1,199,999 counts | 1,199,999 counts |
| Measurement functions | DC voltage, AC voltage, DC current, AC current, DC resistance, capacitance, frequency, continuity, diode, temperature | ||
| Display modes | Direct reading, histogram, bar graph, trend chart | ||
| Measurement speed | Up to 1,000 readings/s | ||
| DC voltage range | 100.000 mV – 1000.00 V | 100.0000 mV – 1000.000 V | 100.0000 mV – 1000.000 V |
| DC voltage accuracy | 0.010 + 0.004% | 0.0035 + 0.0005% | 0.0075 + 0.0005% |
| True RMS AC voltage range | 100.000 mV – 750.000 V | ||
| AC voltage accuracy, 10 Hz – 20 kHz | 0.09 + 0.03% | 0.06 + 0.03% | 0.09 + 0.03% |
| AC voltage accuracy, 3 – 5 Hz | 1.00 + 0.03% | ||
| AC voltage accuracy, 100 – 300 kHz | 4.00 + 0.50% | ||
| DC resistance accuracy (10 Ω – 100 MΩ) | 0.030 + 0.004% | 0.010 + 0.001% | 0.014 + 0.001% |
| DC current accuracy, 100 µA – 10 mA | 0.050 + 0.008% | 0.050 + 0.006% | 0.050 + 0.005% |
| DC current accuracy, 1 A / 3 A | 0.100 + 0.004% / 0.200 + 0.020% | ||
| DC current accuracy, 10 A | 0.250 + 0.004% | 0.120 + 0.010% | 0.120 + 0.010% |
| AC current accuracy, 100 µA – 1 A (3 Hz – 10 kHz) | 0.10 + 0.04% | ||
| AC current accuracy, 3 A / 10 A (3 Hz – 10 kHz) | 0.23 + 0.04% / 0.15 + 0.04% | ||
| Frequency accuracy, 100 Hz – 300 kHz | 0.012% | 0.010% | 0.012% |
| Diode / continuity | 5 V range / 1 kΩ range, test current 1 mA | ||
| Diode / continuity accuracy | 0.1 + 0.02% | 0.010 + 0.030% | 0.1 + 0.02% |
| Capacitance | 1.0000 nF – 10.000 mF; 10.000 nF – 1.0000 mF: 0.5 + 0.1% | ||
| Temperature | PT100 (DIN/IEC 751): probe accuracy ±0.05 °C; 5 kΩ thermistor: probe accuracy ±0.10 °C | ||
| Math functions | Reset, Min/Max/Average/Standard deviation, dB, dBm | ||
| Trigger mode | Local: Auto/Single/Ext · Remote: Immediate/Bus/Ext | ||
| Limit measurement | HI, LO and IN (PASS), with sound beep | ||
| Data logging | Up to 10,000 readings | ||
| Input terminals | Front only | Front & rear | Front only |
| Operating temperature / humidity | 0 – 40 °C, ≤ 90% RH | ||
| Power supply | 99–121 V or 198–242 V AC, auto-selected, 47–63 Hz, 80 VA max. | ||
| Dimensions (rack / bench) | 215 × 88 × 300 mm / 235 × 105 × 320 mm | ||
| Weight | 2.7 kg | ||
Source: TH1953 / TH1963 / TH1963A Series Datasheet, Rev. 2026-09. Specifications subject to change without notice.
Interfaces
- RS232standard
- USB Hoststandard
- USB Devicestandard
- LANstandard
- GPIBoptional (shares slot with HANDLER)
- HANDLERoptional (shares slot with GPIB)
Accessories
Standard accessories
| Item | Description |
|---|---|
| — | EU power cable (Schuko, CEE 7/7) |
| TH26017 | USB interface cable |
| TH26036 | Bench multimeter test lead set |
Optional test fixtures
| Item | Description |
|---|---|
| — | GPIB interface board |
| — | HANDLER interface board |
GPIB and HANDLER share a single rear-panel slot; one or the other can be fitted, not both.
Photos of fixtures and accessories: click an item code above, or browse all accessories.
Downloads
Interested in the TH1953 / TH1963 / TH1963A Series?
Send us the model, quantity and your application. Our team in Ismaning will answer by email.