TH2515
Micro-Ohm DC Resistance Meter, 0.1 µΩ – 110 MΩ
A touch-screen DC low-resistance meter combining 0.01% basic accuracy with 0.1 µΩ resolution for relay-contact, welding-hole, interconnect and conductor resistance testing. Offset voltage compensation and thermal-EMF elimination keep readings free of contact potential effects, while the HANDLER interface and 10-bin comparator make it equally suited to bench use and production-line sorting.
Technical question about this series? Ask our team.
Key features
- 0.01% maximum accuracy, minimum resolution 0.1 µΩ
- 5½-digit reading, resistance range 0.1 µΩ – 110 MΩ
- 4.3" 24-bit colour touchscreen, 480 × 272
- Multiple test-function combinations: R, LPR, T
- Low-voltage test mode (open voltage ≤ 60 mV) protects the DUT
- Temperature compensation (TC) and temperature conversion (Δt)
- Offset voltage compensation (OVC) and user self-correction (0 ADJ)
- 10-bin comparator with CpK and Cp statistics
- Maximum sampling rate 100 samples/s, 4-terminal Kelvin connection
- RS232, USB Host/Device, LAN standard; HANDLER for inline operation; GPIB optional
Model
TH2515
- Basic accuracy
- 0.01%
- Resolution
- 0.1 µΩ
- Measurement range
- 0.1 µΩ – 110 MΩ
- Fastest test speed
- 7 ms
Accuracy and resolution figures above are the instrument's best-case values; see the range table for accuracy by resistance range.
Resistance range and accuracy
| Range | Test current | Resolution | Basic accuracy ±(%Rd + %Fs) | Open-circuit voltage |
|---|---|---|---|---|
| 20 mΩ | 1 A ±5% | 0.1 µΩ | 0.25 + 0.001% | 5 V max |
| 200 mΩ (1 A) | 1 A ±5% | 1 µΩ | 0.25 + 0.001% | 5 V max |
| 200 mΩ (100 mA) | 100 mA ±5% | 1 µΩ | 0.35 + 0.001% | 2.6 V max |
| 2 Ω | 100 mA ±5% | 10 µΩ | 0.035 + 0.001% | 2.6 V max |
| 20 Ω | 10 mA ±5% | 100 µΩ | 0.025 + 0.001% | 2.6 V max |
| 200 Ω | 10 mA ±5% | 1 mΩ | 0.0100 + 0.0010% | 2.6 V max |
| 2 kΩ | 1 mA ±5% | 10 mΩ | 0.0100 + 0.0010% | 2.6 V max |
| 20 kΩ | 100 µA ±5% | 100 mΩ | 0.0100 + 0.0005% | 2.6 V max |
| 100 kΩ | 100 µA ±5% | 1 Ω | 0.0100 + 0.0030% | 13 V max |
| 1 MΩ | 10 µA ±5% | 10 Ω | 0.0200 + 0.0030% | 13 V max |
| 10 MΩ | 1 µA ±5% | 100 Ω | 0.1000 + 0.0060% | 13 V max |
| 100 MΩ | 100 nA ±5% | 1 kΩ | 1.0000 + 0.1000% | 13 V max |
1-year accuracy shown for OVC ON, SLOW2 speed, 23 ± 5 °C, ≤ 80% RH (the instrument's best-case setting). OVC OFF and MED/SLOW1/FAST speeds have looser accuracy; see the Operation Manual for the full table. Outside 18–28 °C a temperature correction factor (K = 1–4×) applies.
Applications
Components and connectors
Resistors, inductors, transformers, motors, relays, circuit solder joints and capacitor riveting points; stranded wire, connectors and switches.
Materials
Heat-sensitive materials such as fuses and thermistor sensors; conductive materials such as metal foil.
New energy and production
EV battery-pack connecting bridges and cell connection resistance; bin-sorted go/no-go inspection on production lines via the HANDLER interface.
Technical specifications
| Parameter | TH2515 |
|---|---|
| Display | 4.3" TFT LCD, 480 × 272, touch screen |
| Reading digits | 5½ digits |
| Resistance measurement time | FAST 7 ms · MED 22 ms · SLOW1 102 ms · SLOW2 402 ms (add 20 ms with display on) |
| Temperature measurement time | 100 ± 10 ms |
| Test terminal | 4-terminal (Kelvin) |
| Averaging | 1 – 255 readings |
| Range switching | AUTO and Manual |
| Trigger mode | Internal, Manual, External, Bus |
| Power-frequency rejection | Selectable, suppresses power-line noise |
| Setting data storage | 30 parameter groups |
| Low-voltage measurement | Open voltage ≤ 60 mV; effective on 2 Ω, 20 Ω, 200 Ω, 2 kΩ ranges |
| Thermal-EMF elimination | Yes |
| Statistics | AVG, MAX, MIN, overall/sample standard deviation, process capability (Cp, CpK) |
| Test-lead contact check | Detects incorrectly connected measurement cables |
| Temperature measurement 1 | Sensor PT500, −99.9–250 °C; 1-yr accuracy ±0.30%Rd ±0.5 °C (≤ 39.9 °C) / ±1.0 °C (40–99.9 °C) |
| Temperature measurement 2 | Analogue input 0–2 V; display range −99.9 – 999.9 °C; 1-yr accuracy ±1%Rd ±3 mV |
| Temperature compensation / conversion | Converts readings to a preset reference temperature; derives temperature rise from before/after resistance values |
| Comparator signal output | HI / IN / LO |
| Comparator limit setup | Absolute high/low limit, or percentage high/low limit + nominal value |
| Sorting | 10 bins, absolute-value or percentage |
| External trigger delay | AUTO (based on range, low-voltage mode, OVC) or MANUAL 0.000–9.999 s |
| Working / storage condition | 0–40 °C, ≤ 80% RH / −10–50 °C, ≤ 90% RH |
| Accuracy guarantee condition | 18–28 °C, ≤ 80% RH |
| Power supply | 99–242 V AC, 47.5–63 Hz, 30 VA |
| Dimensions (rack / bench) | 215 × 87 × 335 mm (net) / 235 × 105 × 360 mm (with foam sheath) |
| Weight | Approx. 3.6 kg |
Source: TH2515 Datasheet, Rev. 2026-09. Specifications subject to change without notice.
Interfaces
- RS232Cstandard
- USB Hoststandard
- USB Devicestandard
- LANstandard (LXI Class C)
- Handlerstandard
- GPIBoptional
Accessories
Standard accessories
| Item | Description |
|---|---|
| — | EU power cable (Schuko, CEE 7/7) |
| TH26050S | Four-terminal Kelvin test cable |
| PT500 | Temperature sensor |
Optional test fixtures
| Item | Description |
|---|---|
| — | GPIB interface module |
| — | High-resistance noise-abatement function for multi-pole connector |
Additional test leads and fixtures are available on request from Techmize or your authorized distributor.
Photos of fixtures and accessories: click an item code above, or browse all accessories.
Downloads
Interested in the TH2515?
Send us the model, quantity and your application. Our team in Ismaning will answer by email.