Low-Resistance (DCR) Meters

TH2516 Series

DC Low-Resistance Meter, 1 µΩ – 2 MΩ

A touchscreen DC resistance meter family for testing relay contacts, interconnects, solder joints and welded connections, with resistance measurement down to 1 µΩ and up to 0.05% basic accuracy. Four-wire Kelvin terminals, offset voltage compensation and thermal-EMF elimination keep readings stable on low-resistance DUTs. Three models span a full 1 µΩ – 2 MΩ flagship down to an economy variant for production testing.

TH2516TH2516ATH2516B CE

Technical question about this series? Ask our team.

Key features

  • 0.05% basic accuracy, minimum resolution down to 1 µΩ
  • 4.3" TFT colour touchscreen, 480 × 272, English UI
  • Four-wire (Kelvin) test terminals with R, LPR and T measurement modes
  • Low-voltage test mode (open-circuit ≤ 40 mV) protects sensitive DUTs
  • Offset voltage compensation (OVC) and thermal-EMF elimination
  • Temperature compensation (TC) and temperature conversion (Δt) on TH2516
  • Built-in comparator with 3-bin sorting (HI/IN/LO), audible pass/fail
  • Statistics: AVG, MAX, MIN, standard deviation, Cp, CpK
  • User self-calibration (0 ADJ), 30 stored setting groups, USB screen capture
  • RS232, USB Host/Device, HANDLER standard; SCPI and MODBUS; LXI Class C compliant

3 models, one platform

TH2516

Full-range flagship with temperature measurement

Range
1 µΩ – 2 MΩ
Basic accuracy
0.05% (typ.)
Max. current
1 A
Temperature meas.
Yes
Request a quote for TH2516

TH2516A

Mid-range, 0.05% accuracy, no temperature option

Range
10 µΩ – 200 kΩ
Basic accuracy
0.05% (typ.)
Max. current
100 mA
Temperature meas.
—
Request a quote for TH2516A

TH2516B

Entry-level for production testing

Range
1 µΩ – 20 kΩ
Basic accuracy
0.1% (typ.)
Max. current
1 A
Temperature meas.
—
Request a quote for TH2516B

TH2516 is the only model with temperature measurement (PT500 sensor input and 0–2 V analog input) and ships with a PT500 probe as standard. Low-voltage test mode and thermal-EMF elimination are available on TH2516 and TH2516A. All three models share the same 4-wire Kelvin terminals, comparator, statistics function and interface set.

Applications

Components

Resistors, inductors, transformers, motors, relays, circuit solder joints and capacitor riveting points.

Cables and connectors

Multi-strand wire, connectors and switches of all types.

Materials and new energy

Heat-sensitive materials (fuses, thermistor sensors) and conductive metal foils; EV battery-pack connecting bridges and cell connection resistance.

Technical specifications

ParameterTH2516TH2516ATH2516B
Display 4.3" colour TFT LCD, 480 × 272, touchscreen
Reading digits 4½ digits
Measurement range 1 µΩ – 2 MΩ10 µΩ – 200 kΩ1 µΩ – 20 kΩ
Basic accuracy ¹ 0.05% (typ.)0.05% (typ.)0.1% (typ.)
Test current up to 1 Aup to 100 mAup to 1 A
Test functions R, LPR, T
Test terminal 4-wire (Kelvin)
Measurement speed FAST 10 ms · MED 25 ms · SLOW1 115 ms · SLOW2 455 ms
Max. sample rate 50 samples/s
Averaging 1 – 255 readings
Range switching Auto, Manual
Trigger mode Internal, Manual, External, Bus
Low-voltage test mode ≤ 40 mV open-circuit; 2 Ω / 20 Ω / 200 Ω / 2 kΩ≤ 40 mV open-circuit; 2 Ω / 20 Ω / 200 Ω / 2 kΩ—
Thermal-EMF elimination ✓✓—
Offset voltage compensation ✓ (OVC)
Zero adjustment ✓ (0 ADJ, user self-calibration)
Temperature sensor input PT500, −99.9 to 250 °C——
Temperature analog input 0–2 V, −99.9 to 999.9 °C——
Temperature compensation / Δt ✓——
Comparator output HI/IN/LO signal output
Comparator beep OFF, IN, HI/LO
Limit setup 3 bins, absolute high/low limit or percentage high/low limit + nominal value
Statistics function AVG, MAX, MIN, overall/sample std. deviation, Cp, CpK
Setting storage 30 groups
Data and screen capture Screenshot and data logging to USB
Compliance CE; LXI Class C
Working / storage condition 0–40 °C, ≤ 80% RH / −10–50 °C, ≤ 90% RH
Power supply 99–121 V AC / 198–242 V AC, 47.5–63 Hz, 30 VA
Dimensions (rack / bench) 215 × 88 × 335 mm / 235 × 105 × 360 mm
Weight 3.6 kg

¹ Accuracy is range-dependent; the lowest range (20 mΩ) and, on TH2516, the highest range (2 MΩ) carry relaxed specifications. Ask Techmize for the full per-range accuracy table.

Source: TH2516 Series Datasheet, Rev. 2026-09. Specifications subject to change without notice.

Interfaces

  • RS232standard
  • USB Hoststandard
  • USB Devicestandard
  • Handlerstandard
  • SCPI / MODBUSstandard

Accessories

Standard accessories

ItemDescription
—EU power cable (Schuko, CEE 7/7)
TH26050SFour-terminal Kelvin test cable
PT500Temperature sensor (TH2516 only)

Contact Techmize or your authorized distributor for test fixtures and cabling for specific DUT geometries.

Photos of fixtures and accessories: click an item code above, or browse all accessories.

Downloads

Interested in the TH2516 Series?

Send us the model, quantity and your application. Our team in Ismaning will answer by email.

Request a quote