TH2852 / TH2853 Series
RF Precision Impedance Analyzer & RF LCR Tester, 1 MHz – 5 GHz
A new generation of RF impedance test instruments based on the RF-IV measurement method, designed to keep high accuracy well into the GHz range. An external test head with automatic electrical-length measurement and calibration extends RF-IV testing up to 5 GHz, for high-precision testing of wafers and PCBs. Each series is available in five frequency-range models, from 500 MHz up to 5 GHz.
Technical question about this series? Ask our team.
Key features
- Test frequency 1 MHz – 5 GHz, five frequency-range models per series
- ±0.3% basic accuracy (typical), RF-IV method for stable accuracy at GHz frequencies
- High speed: measurement time as fast as 0.4 ms per point
- 10.1" capacitive touch screen, 1280 × 800 resolution
- Four-parameter simultaneous measurement; 8 quick-select parameter-set keys
- Point, list and curve-sweep measurement, with automatic 10-bin sorting
- 1601-point multi-parameter list sweep (frequency, AC level, DC bias)
- 4-channel graphical sweep, up to 4 curves per channel, 14 split-screen layouts
- Segmented frequency sweep, e.g. for crystal-resonator testing
- External test head with automatic electrical-length calibration for wafer/PCB testing up to 5 GHz
10 models, one platform
| Model | Series | Test frequency | Quote |
|---|---|---|---|
| TH2852-050Precision impedance analyzer | TH2852 | 1 MHz – 500 MHz | Quote |
| TH2852-100Precision impedance analyzer | TH2852 | 1 MHz – 1 GHz | Quote |
| TH2852-200Precision impedance analyzer | TH2852 | 1 MHz – 2 GHz | Quote |
| TH2852-350Precision impedance analyzer | TH2852 | 1 MHz – 3.5 GHz | Quote |
| TH2852-500Precision impedance analyzer | TH2852 | 1 MHz – 5 GHz | Quote |
| TH2853-050RF LCR tester | TH2853 | 1 MHz – 500 MHz | Quote |
| TH2853-100RF LCR tester | TH2853 | 1 MHz – 1 GHz | Quote |
| TH2853-200RF LCR tester | TH2853 | 1 MHz – 2 GHz | Quote |
| TH2853-350RF LCR tester | TH2853 | 1 MHz – 3.5 GHz | Quote |
| TH2853-500RF LCR tester | TH2853 | 1 MHz – 5 GHz | Quote |
Ten individually orderable models: two series, five frequency ranges each. The suffix (-050 to -500) sets only the maximum test frequency. Basic accuracy, AC/DC source ranges, display and interfaces are identical between the two series at the same suffix.
TH2852 vs. TH2853
| Function | TH2852 Series | TH2853 Series |
|---|---|---|
| Minimum frequency resolution | 1 mHz | 1 kHz |
| Test functions | Point, list, curve sweep, Smith chart | Point, list, curve sweep (fixed-grid display) |
| Curve-sweep cursor functions | ✓ | — |
| Curve-sweep statistical analysis | ✓ | — |
| Curve-sweep data storage | ✓ | — |
| Equivalent-circuit analysis | ✓ | — |
| Smith chart | ✓ | — |
| Permeability / dielectric-constant test | Optional | — |
TH2853 is a simplified, cost-effective platform for production and incoming-inspection testing where cursor analysis, curve statistics and equivalent-circuit fitting are not required. Dielectric-constant and permeability analysis on TH2852 use the TH26089 / TH26088 fixtures.
Applications
Passive and semiconductor components
Capacitors, inductors, PCB inductors, magnetic cores, resistors, wireless-charging and NFC/RFID coils, voice coils, motor coils, crystal oscillators, piezoelectric devices, transformers and chip components; parasitic-parameter analysis of LED driver ICs, varactor diodes and transistors.
Other components
Impedance evaluation of printed circuit boards, relays, switches, cables and batteries.
Materials analysis (optional)
Dielectric constant and loss tangent of plastics, ceramics and PCB materials; permeability and loss tangent of ferrites and amorphous magnetic materials; C-V characteristics of semiconductor materials and liquid-crystal cells.
Technical specifications
| Parameter | -050 | -100 | -200 | -350 | -500 |
|---|---|---|---|---|---|
| Test frequency | 1 MHz – 500 MHz | 1 MHz – 1 GHz | 1 MHz – 2 GHz | 1 MHz – 3.5 GHz | 1 MHz – 5 GHz |
| Display | 10.1" TFT LCD, 1280 × 800, capacitive touch screen | ||||
| Max. frequency resolution | 1 mHz (TH2852 series), 1 kHz (TH2853 series) | ||||
| AC measurement parameters | |Z|, |Y|, θ, R, X, G, B, L, C, D, Q, |Γ|, Γx, Γy, θΓ, Vac, Iac, impedance circle, admittance circle, Smith chart¹, dielectric/magnetic material measurement (optional, TH2852 only) | ||||
| DC measurement parameters | VDC, IDC, RDC | ||||
| AC test level | Voltage −41 dBm to +1 dBm; current 80 µA – 10.04 mArms | ||||
| DC bias source | Voltage 0–±40 V; current 0–±100 mA | ||||
| Measurement speed (5 levels) | 0: 0.4 ms, 1: 4 ms, 2: 26 ms, 3: 65 ms, 4: 130 ms, 5: 650 ms | ||||
| Basic accuracy | ±0.3% (typical value) | ||||
| Impedance test range | 100 mΩ – 60 kΩ | ||||
| Measurement display range | Cs, Cp: ±9.99999 aF – 9.99999 EF; Ls, Lp: ±1.00000 aH – 99.9999 EH; D: ±0.00001 – 9.99999; Q: ±0.01 – 99999.9; R, X, Z: ±1.00000 aΩ – 999.999 EΩ; G, B, Y: ±1.00000 aS – 99.9999 ES | ||||
| Sorting / binning | LCR mode: 10-grade sorting; scan mode: independent sorting per curve | ||||
| Warm-up time | 30 minutes | ||||
| Input voltage | 100–120 V AC / 198–242 V AC selectable, 47–63 Hz | ||||
| Power consumption | Max. 150 VA | ||||
| Dimensions (W × H × D) | 428 × 220 × 280 mm | ||||
| Weight | 14 kg | ||||
¹ Smith chart and dielectric/magnetic material measurement are TH2852-series capabilities.
Source: TH2852 / TH2853 Series Datasheet, Rev. 2026-09. Specifications subject to change without notice.
Interfaces
- RS232Cstandard
- USB Hoststandard
- USB Devicestandard
- LANstandard
- GPIBstandard
- Handlerstandard
- VGAstandard
- HDMIstandard
Accessories
Standard accessories
| Item | Description |
|---|---|
| — | EU power cable (Schuko, CEE 7/7) |
| TH2852-1 | 0 m test head (standard with TH2852 models) |
| TH2852-1A | 1 m test head (standard with TH2853 models) |
Optional test fixtures
| Item | Description |
|---|---|
| TH2852-1 | 0 m test head |
| TH2852-1A | 1 m test head |
| TH2852-1B | 2 m test head |
| — | APC7mm to SMA female adapter |
| — | TH2852 calibration kit (open/short/load/capacitance standards) |
| TH2852-3A | 0805 parallel-electrode fixture |
| TH2852-3B | 0603 parallel-electrode fixture |
| TH2852-3C | 0402 parallel-electrode fixture |
| TH2852-3D | 0201 parallel-electrode fixture |
| TH2852-4 | Low-frequency through-hole / parallel-electrode fixture |
| TH2852-5 | Large parallel-electrode fixture |
| TH2852-6 | Small bottom-electrode fixture |
| TH26088 | Permeability test fixture |
| TH26089 | Dielectric-constant test fixture |
More options in preparation. Contact Techmize or your authorized distributor for details.
Photos of fixtures and accessories: click an item code above, or browse all accessories.
Downloads
Interested in the TH2852 / TH2853 Series?
Send us the model, quantity and your application. Our team in Ismaning will answer by email.