Hipot & Insulation Resistance Testers
TH9130S Series
Multi-Channel High-Power Withstand-Voltage & Ground-Bond Tester
A multi-channel derivative of the TH9130 platform, built for production-line safety testing where several DUTs or test points are scanned in sequence. Combines AC/DC withstand voltage, insulation resistance, ground bond and continuity testing with user-selectable 4- or 8-channel rear-panel scanning modules and hardware contact check (CK) for reliable multi-point testing.
Technical question about this series? Ask our team.
Key features
- 7-inch capacitive touchscreen, 800 × 480, English UI
- Five-in-one: ACW, DCW, insulation resistance, ground bond and continuity, with multi-channel scanning
- 500 VA AC withstand-voltage output (TH9130S); insulation resistance test up to 6 kV / 50 GΩ
- Ground bond test up to 40 A; can run simultaneously with withstand voltage
- Hardware contact check (CK) confirms probe contact before applying high voltage
- Rear-panel scanning module: 4 or 8 HV channels, 4 or 8 ground-bond channels (2-wire or 4-wire)
- Open/short-circuit detection (OSC) and arc detection (ARC)
- Single-screen step view plus list mode with all step results and pass/fail
- Shock-protection cut-off, lockable panel, pass/fail alarm
- Standard SCPI; storage for 100 files, 50 steps per file
2 models, one platform
TH9130S
- AC withstand
- 5.0 kV / 100 mA / 500 VA
- DC withstand
- 6.0 kV / 20 mA / 120 VA
- Synchronous ACW + GB
- 5 kVac/30 mAac & 30 Aac/150 mΩ
- Weight
- 40 kg
TH9131S
- AC withstand
- 5.0 kV / 40 mA / 200 VA
- DC withstand
- 6.0 kV / 20 mA / 120 VA
- Synchronous ACW + GB
- 5 kVac/30 mAac & 30 Aac/150 mΩ
- Weight
- 38 kg
A rear-panel scanning module must be selected when ordering; the base instrument does not ship without one. The module is common to TH9130S and TH9131S.
Scanning module options
| Feature | -004 | -001 | -002 | -003 |
|---|---|---|---|---|
| HV channels (ACW / DCW / IR) | 4 | 8 | 8 | — |
| Hardware contact check (CK) | ✓ | ✓ | ✓ | — |
| Ground-bond (GB) channels | 4 | 8 | — | 8 |
| GB wiring | Four-wire | Two-wire | — | Four-wire |
| Continuity | ✓ | ✓ | ✓ | ✓ |
Module -004 offers fewer channels with four-wire (Kelvin) ground-bond wiring; module -001 maximises channel count with two-wire ground bond; module -002 is high-voltage scanning only; module -003 is ground-bond scanning only (no high-voltage channels, no CK).
Applications
Household appliances and lighting
Electrical safety testing on the production line; comprehensive lighting-appliance safety testing.
Motors and high-power equipment
Comprehensive electric-motor safety testing; withstand-voltage and ground-bond testing of high-power appliances.
Electronic components
Multi-point production-line component safety testing.
Technical specifications
| Parameter | TH9130S | TH9131S |
|---|---|---|
| AC voltage range | 0.05 – 5.0 kV, 50/60 Hz ±0.1% sine wave | |
| AC output power | 500 VA (5.0 kV / 100 mA) | 200 VA (5.0 kV / 40 mA) |
| DC voltage range | 0.05 – 6.0 kV | |
| DC output power | 120 VA (6.0 kV / 20 mA) | |
| Voltage accuracy | ±(1% set value + 5 V); resolution 2 V | |
| AC current range | 0.001 – 120 mA (V ≤ 4 kV) / 100 mA (V > 4 kV) | 0.001 – 40 mA |
| DC current range | 0.0001 – 20 mA | |
| Test / rise / fall / wait time | 0.3–999 s / 0.1–999 s / 0.1–999 s / 0.1–999 s (0 = continuous / off) | |
| Arc detection (AC / DC) | 1.0–20.0 mA / 1.0–10.0 mA | 1.0–20.0 mA / — |
| Max. short-circuit current (AC) | 200 mA | 80 mA |
| IR output voltage | DC 0.05 – 6.0 kV, resolution 2 V, accuracy ±(1% reading + 5 V) | |
| IR resistance range | 0.05 MΩ – 50.0 GΩ | |
| IR accuracy (V ≥ 500 V) | 1 MΩ–1 GΩ: ±(3% reading + 1 MΩ); 1–10 GΩ: ±(7% reading + 0.2 GΩ); 10–50 GΩ: ±(10% reading + 0.5 GΩ) | |
| IR accuracy (V < 500 V) | 1 MΩ – 1 GΩ: ±(5% reading + 100 V/Vs × 10 MΩ) | |
| Ground bond output current | 1.00 – 40.00 A, ±(2% set value + 2 digits) | |
| Ground bond test voltage | 3.00 – 8.00 V, 50/60 Hz ±0.1% | |
| Ground bond resistance range | 0–600 mΩ (1–10 A), 0–200 mΩ (10–30 A), 0–150 mΩ (30–40 A); accuracy ±(2–3% reading + 2–3 digits) | |
| Synchronous ACW + GB test | 5 kVac/30 mAac & 30 Aac/150 mΩ | |
| Continuity | 0.0001–0.1 A / 0.00–10,000 Ω; accuracy ±(1% reading + 3–10 digits) | |
| OSC detection | 0.001–40 nF; open 10–100%, short 100–500% | |
| HV channels | 4 (module -004) or 8 (modules -001 / -002); none on module -003 | |
| Ground-bond channels | 4 (module -004, four-wire), 8 (module -001, two-wire), 8 (module -003, four-wire); none on module -002 | |
| Contact check (CK) | Standard on modules -004 / -001 / -002; not available on module -003 | |
| Shock protection | 0.5 mA ±0.25 mA, selectable on/off; interlock pin must be pulled low to enable HV output | |
| Alarm / lock | Pass: short tone + green light. Fail: long tone + red light. Lock-on-fail, panel key-lock | |
| Storage | 100 files, 50 steps per file | |
| Operating / storage environment | 0–45 °C, 20–90% RH / −10–55 °C, < 80% RH | |
| Power supply / power | 100–240 V AC, 47–63 Hz; no-load < 100 W, rated 1200 W | |
| Dimensions (W × H × D) | 430 × 132 × 550 mm | |
| Weight | 40 kg (all modules) | 38 kg (all modules) |
Source: TH9130S / TH9131S Series Datasheet, Rev. 2026-09. Specifications subject to change without notice.
Interfaces
- RS232standard
- USB Hoststandard
- USB Devicestandard
- LANstandard
- Handlerstandard
- SCPIstandard
- GPIBoptional
Accessories