TH2883 Series
Impulse Winding Tester, up to 10 kV
A high-voltage impulse winding tester family for non-destructive insulation testing of motor, transformer and wound-inductor coils. The TH2883 series compares the damped-oscillation waveform of the unit under test with a stored reference waveform to reveal turn-to-turn and layer-to-layer short circuits, winding-count differences and core-material variation. Impulse output spans 30 V to 10 kV across three single-channel models.
Technical question about this series? Ask our team.
Key features
- 30 V – 10 kV impulse voltage across the series; minimum testable inductance 1 µH / 10 µH / 20 µH depending on model
- Up to 200 Msps waveform sampling, 8-bit resolution, 6k-byte memory depth
- Corona-discharge extraction algorithm with high-bandwidth analog acquisition
- Four comparison methods: area, differential area, corona discharge, differential phase
- 7" 65k-colour TFT display with English UI
- Step-voltage destructive test (BDV) mode to help establish a safe test-voltage margin
- Demagnetizing-pulse application keeps repeated test results consistent
- Up to 32-pulse sample averaging; 20 internally stored standard-waveform files
- Scannable to 10 channels with the optional TH28831 interturn-insulation scanner
- LAN remote control; SCPI supported
3 models, one platform
TH2883-1
Low-inductance coils, e.g. switching-power-supply windings
- Impulse voltage
- 30 V – 1.2 kV
- Min. inductance
- ≥ 1 µH
- Impulse energy
- ≤ 0.02 J
- Channels
- 1
TH2883-5
Standard model for general coil testing
- Impulse voltage
- 100 V – 5 kV
- Min. inductance
- ≥ 10 µH
- Impulse energy
- ≤ 0.25 J
- Channels
- 1
TH2883-10
Higher-voltage interturn insulation testing
- Impulse voltage
- 500 V – 10 kV
- Min. inductance
- ≥ 20 µH
- Impulse energy
- ≤ 0.5 J
- Channels
- 1
TH2883-1's 1.2 kV ceiling applies to sufficiently high winding inductance; the maximum output voltage falls at very low inductance values. Consult Techmize for the voltage-vs-inductance curve if testing near the 1 µH minimum. The 8- and 4-channel models are listed as the TH2883S Series.
Applications
Motor and generator windings
Turn-to-turn and layer-to-layer insulation testing on motor and generator windings.
Transformer windings
Interturn and interlayer insulation and short-circuit detection on transformer windings.
Wound inductors and coils
Insulation and short-circuit testing of wound inductors, with PASS/FAIL sorting for production lines.
Industries and application examples
Technical specifications
| Parameter | TH2883-1 | TH2883-5 | TH2883-10 |
|---|---|---|---|
| Channels | 1 | ||
| Impulse voltage | 30 V – 1.2 kV, 5 V step | 100 V – 5 kV, 10 V step | 500 V – 10 kV, 20 V step |
| Output voltage setting accuracy | ±(5% of setting + 5 V) | ±(5% of setting + 15 V) | ±(5% of setting + 25 V) |
| Voltage readback accuracy | ±(5% of reading + 5 V) | ±(5% of reading + 15 V) | ±(5% of reading + 25 V) |
| Minimum testable inductance | ≥ 1 µH | ≥ 10 µH | ≥ 20 µH |
| Impulse energy | ≤ 0.02 J | ≤ 0.25 J | ≤ 0.5 J |
| Measurement speed | 6 meas/s (167 ms) | 6 meas/s (167 ms) | 3 meas/s at 10 kV (333 ms) |
| Applied pulses | Up to 32 per measurement (sample averaging) | ||
| Input impedance | 5 MΩ | ||
| Display | 7", 800 × 480-dot, 65k-colour TFT; waveform display area 600 × 256 | ||
| Waveform acquisition | Sampling rate up to 200 Msps, 8 adjustable levels; 8-bit resolution; 6k-byte memory depth; sample averaging 1–32 | ||
| Comparison methods | Area, differential area, corona discharge, differential phase | ||
| Waveform measurement | Voltage, time, frequency | ||
| Trigger mode | Manual (incl. foot switch), external, internal, bus | ||
| Detection output | OK/NG display, LED indicator, buzzer alarm | ||
| Measurement statistics | Pass/fail count and pass-rate statistics per comparison method | ||
| Memory | 20 internal standard-waveform/setup files; USB flash drive for external storage | ||
| Power supply | 110 V / 220 V AC ±10%, 50/60 Hz ±5%; ≤ 200 VA | ||
| Working environment | 0–40 °C, ≤ 75% R.H. | ||
| Safety and EMC | IEC 61010-1:2001, IEC 61326-2-1:2005 | ||
| Dimensions / weight | 400 (W) × 132 (H) × 420 (D) mm, 15 kg | ||
Source: TH2883 Series Datasheet, Rev. 2026-09. Specifications subject to change without notice.
Interfaces
- RS232Cstandard
- USB Hoststandard
- USB Devicestandard
- LANstandard (web-browser access and remote control)
- Handlerstandard
- SCPIstandard (MODBUS not supported)
- Foot switch inputstandard
Accessories
Standard accessories
| Item | Description |
|---|---|
| — | EU power cable (Schuko, CEE 7/7) |
| TH2883-01 | High-voltage test cable |
| TH2881-001 | Foot switch |
Photos of fixtures and accessories: click an item code above, or browse all accessories.
Downloads
Interested in the TH2883 Series?
Send us the model, quantity and your application. Our team in Ismaning will answer by email.