Hipot & Insulation Resistance Testers
TH9110 Series
AC/DC Withstand Voltage & Insulation Resistance Tester, up to 500 VA
A digitally controlled hipot and insulation-resistance test platform for transformers, motors, windings, passive components and finished electrical products. Closed-loop digital output regulation keeps the test voltage stable regardless of mains fluctuation, while ARC detection, open/short (OSC) contact checking and breakdown-voltage ramp testing improve safety-test reliability and throughput. Five models differ in output power, current range and the availability of high-voltage floating output or CK pin/contact detection.
Technical question about this series? Ask our team.
Key features
- High power output, up to 5 kV / 100 mA / 500 VA AC/DC (TH9110 / TH9110A / TH9112A)
- Digitally controlled, closed-loop output regulation, unaffected by mains fluctuation
- 7" 800 × 480 TFT-LCD display, English UI, large-character PASS/FAIL readout
- High-voltage floating output (TH9110 / TH9111), meeting EN 50191 safety requirements
- Electric-shock protection, selectable on all five models
- ARC detection for early insulation-degradation and arc-discharge detection
- Contact check (OSC) flags open or shorted test leads before they cause false results
- CK pin/contact detection (TH9112A) confirms DUT connection before the HV test
- Breakdown-voltage ramp test mode for dielectric-strength analysis
- One-key screen capture; rear-panel HV output; 100 files, up to 50 steps per file
5 models, one platform
| Model | AC power | Floating output | Leakage | Weight | Quote |
|---|---|---|---|---|---|
| TH9110High power, floating output | 500 VA | Yes | < 3.5 mA | 21 kg | Quote |
| TH9110AHigh power, standard output | 500 VA | — | — | 21 kg | Quote |
| TH9111Compact power, floating output | 200 VA | Yes | < 3.5 mA | 20 kg | Quote |
| TH9111ACompact power, standard output | 200 VA | — | — | 20 kg | Quote |
| TH9112AHigh current, CK pin detection | 500 VA | — | — | 21 kg | Quote |
TH9110/TH9110A/TH9111/TH9111A cover 0.05–5.00 kV AC / 0.05–6.00 kV DC withstand voltage and 0.05–5.00 kV insulation resistance (0.1 MΩ – 50.0 GΩ); TH9110/TH9111 add high-voltage floating output. TH9112A is the high-current variant: same voltage ranges, 500 VA output in AC and DC modes, insulation resistance to 100 GΩ, and CK pin/contact detection instead of floating output.
Applications
Windings and high-power products
Transformers, generators/motors and other winding products, such as motor stators and rotors with high parasitic capacitance, requiring high-power withstand-voltage testing.
Components and electrical products
Capacitors, coils, cores, choke coils and filters; household appliances, information and audio-visual products, electric heating appliances and lighting equipment; wire, non-woven fabric and insulation materials.
Automated and specialised testing
Production-line automation via rear-panel HV output; automated test systems; new-energy vehicle components; medical equipment.
Industries and application examples
Technical specifications
| Parameter | TH9110 | TH9110A | TH9111 | TH9111A | TH9112A |
|---|---|---|---|---|---|
| AC output voltage | 0.05 – 5.00 kV, 50/60 Hz ±0.1% sine wave | ||||
| Output power | 500 VA | 500 VA | 200 VA | 200 VA | 500 VA |
| DC output voltage | 0.05 – 6.00 kV | ||||
| DC output ripple | — | — | — | — | −2% / +3% of avg. (5 kV / 100 mA resistive load) |
| Load variance | ±(1% setting + 0.2% F.S.), at rated power | ||||
| Voltage resolution / accuracy | 2 V / ±(1% setting + 0.1% F.S.) | ||||
| AC current range | 0.001–120 mA (≤ 4 kV), 0.001–100 mA (> 4 kV) | 0.001–120 mA (≤ 4 kV), 0.001–100 mA (> 4 kV) | 0.001–40 mA | 0.001–40 mA | 0.001–120 mA (≤ 4 kV), 0.001–100 mA (> 4 kV) |
| AC current resolution / accuracy | 1 µA / ±(1% reading + 0.5% F.S.) | ||||
| DC current range | 0.0001–25 mA (≥ 1.5 kV), 0.0001–20 mA (< 1.5 kV) | 0.0001–25 mA (≥ 1.5 kV), 0.0001–20 mA (< 1.5 kV) | 0.0001–20 mA (< 1.5 kV) | 0.0001–20 mA (< 1.5 kV) | 0.0001–100 mA (≤ 5 kV), 0.0001–90 mA (5–5.5 kV), 0.0001–80 mA (> 5.5 kV) |
| DC current resolution / accuracy | 0.1 µA / ±(1% reading + 0.5% F.S.) | ||||
| Max. short-circuit current | 200 mA (AC test only) | 200 mA (AC test only) | 80 mA (AC test only) | 80 mA (AC test only) | 200 mA |
| IR output voltage | DC 0.05–5.00 kV, 2 V res., ±(1% reading + 0.1% F.S.) | DC 0.05–5.00 kV, 2 V res., ±(1% reading + 0.1% F.S.) | DC 0.05–5.00 kV, 2 V res., ±(1% reading + 0.1% F.S.) | DC 0.05–5.00 kV, 2 V res., ±(1% reading + 0.1% F.S.) | DC 0.05–6.00 kV, 2 V res., ±(1% reading + 0.1% F.S.) |
| IR resistance range | 0.1 MΩ – 50.0 GΩ | 0.1 MΩ – 50.0 GΩ | 0.1 MΩ – 50.0 GΩ | 0.1 MΩ – 50.0 GΩ | 0.1 MΩ – 100.0 GΩ |
| IR accuracy ≥ 1 kV | 1–1000 MΩ: ±(3% rdg + 0.1% F.S.); 1–10 GΩ: ±(7% rdg + 2% F.S.); 10–50 GΩ: ±(10% rdg + 1% F.S.) | 1–1000 MΩ: ±(3% rdg + 0.1% F.S.); 1–10 GΩ: ±(7% rdg + 2% F.S.); 10–50 GΩ: ±(10% rdg + 1% F.S.) | 1–1000 MΩ: ±(3% rdg + 0.1% F.S.); 1–10 GΩ: ±(7% rdg + 2% F.S.); 10–50 GΩ: ±(10% rdg + 1% F.S.) | 1–1000 MΩ: ±(3% rdg + 0.1% F.S.); 1–10 GΩ: ±(7% rdg + 2% F.S.); 10–50 GΩ: ±(10% rdg + 1% F.S.) | 1–1000 MΩ: ±(3% rdg + 1 MΩ); 1–10 GΩ: ±(7% rdg + 0.2 GΩ); 10–100 GΩ: ±(10% rdg + 0.5 GΩ) |
| IR accuracy < 500 V | 0.1–1000 MΩ: ±(5% rdg + 2% F.S.) | 0.1–1000 MΩ: ±(5% rdg + 2% F.S.) | 0.1–1000 MΩ: ±(5% rdg + 2% F.S.) | 0.1–1000 MΩ: ±(5% rdg + 2% F.S.) | 0.1–1000 MΩ: ±(5% rdg + 100 V/Vs × 10 MΩ) |
| ARC detection level | AC 1.0 – 20.0 mA / DC 1.0 – 10.0 mA | ||||
| OSC sample capacitance / thresholds | 0.001 – 40 nF; open 10%–100%, short 100%–500% (test signal 600 Hz, 1 s) | ||||
| CK pin/contact detection | — | — | — | — | Test voltage 0.1–0.5 kV, current lower limit 0.001–5 mA |
| Breakdown-voltage test | AC / DC; start 0.05 kV to end 5 kV; step mode (2–999 steps) or voltage-step mode (5–500 V); 0.3–999 s per step | ||||
| Test time | 0.3 – 999 s (0 = continuous) | ||||
| Rise / fall time | 0.1 – 999.9 s (0 = off) | ||||
| Waiting time (DC / IR) | 0.1 – 999 s (0 = off) | ||||
| Time accuracy | ±(1% + 0.1 s) | ||||
| High-voltage floating output | ACW/DCW/IR, leakage < 3.5 mA | — | ACW/DCW/IR, leakage < 3.5 mA | — | — |
| Electric-shock protection | 0.5 mA ±0.25 mA, selectable on/off | ||||
| Interlock / panel protection | Interlock pin must be pulled low to enable HV output; key lock, password | ||||
| Internal storage | 100 files, up to 50 steps per file | ||||
| Operating / storage temperature | 0–45 °C, 20–90% RH / −10–55 °C, < 80% RH | ||||
| Power supply / consumption | 100 – 240 V AC, 47 – 63 Hz; idle < 100 W, rated 1000 W | ||||
| Dimensions (W × H × D) | 430 × 132 × 500 mm | ||||
| Weight | 21 kg | 21 kg | 20 kg | 20 kg | 21 kg |
Source: TH9110 Series Datasheet, Rev. 2026-09. Specifications subject to change without notice.
Interfaces
- RS232standard
- USB Hoststandard
- USB Devicestandard
- LANstandard
- Handlerstandard
- SCPIstandard
- GPIBoptional
Accessories
More accessories are available on request from Techmize or your authorized distributor.
Photos of fixtures and accessories: click an item code above, or browse all accessories.
Downloads
Interested in the TH9110 Series?
Send us the model, quantity and your application. Our team in Ismaning will answer by email.