Hipot & Insulation Resistance Testers
TH9320-S Series
4/8-Channel Scanning AC/DC Withstand Voltage Tester
The TH9320-S series adds built-in high-voltage scanning to the TH9320 platform, so a single instrument can sequence AC/DC withstand-voltage, insulation-resistance and contact-check tests across multiple points on a DUT, or across multiple DUTs, without external switching hardware. It shares the display, control panel, test performance and operation manual of the TH9320 series.
Technical question about this series? Ask our team.
Key features
- AC withstand voltage 0.05–5.00 kV, up to 20.00 mA, 50/60 Hz selectable
- DC withstand voltage 0.05–6.00 kV, up to 10.00 mA
- Insulation resistance test 0.10–1.00 kV, 0.1 MΩ – 10.0 GΩ
- Built-in 4-channel (TH9320-S4) or 8-channel (TH9320-S8) high-voltage scanning
- Programmable contact-check steps verify DUT connection before withstand-voltage testing
- 480 × 272-dot TFT-LCD colour display, English UI
- Rapid discharge and arc detection (ARC)
- Ground-leakage (GFI) shock protection
- 20 stored test files, 20 test steps per file
- Standard RS232, USB Host, USB Device, Handler, Signal, PLC; SCPI and MODBUS
2 models, one platform
TH9320-S4
4-channel scanning
- Scan channels
- 4
- AC withstand
- 5 kV / 20 mA
- DC withstand
- 6 kV / 10 mA
- Contact check
- Yes
TH9320-S8
8-channel scanning
- Scan channels
- 8
- AC withstand
- 5 kV / 20 mA
- DC withstand
- 6 kV / 10 mA
- Contact check
- Yes
TH9320-S4 and TH9320-S8 share identical test performance, chassis, display and interfaces; they differ only in the number of built-in scan channels. Each channel is sequenced automatically as part of a stored test file.
Applications
Test and production systems
Automated test systems, electronic components, electrical equipment safety testing.
Consumer and household
Household appliances, electric heating appliances, lighting products.
Industrial and e-mobility
Transformers and motors, new-energy-vehicle components, medical devices.
Industries and application examples
Technical specifications
| Parameter | TH9320-S4 | TH9320-S8 |
|---|---|---|
| Display | 480 × 272-dot TFT-LCD, colour | |
| Scan interface | 4 channels | 8 channels |
| AC output voltage | 0.05 – 5.00 kV, ±(2% reading + 5 V); 50 Hz / 60 Hz selectable | |
| AC current test range | 1 µA – 20.00 mA, ±(2% reading + 5 digits) | |
| DC output voltage | 0.05 – 6.00 kV, ±(2% reading + 10 V) | |
| DC current test range | 0.1 µA – 10.00 mA, ±(2% reading + 5 digits) | |
| DC discharge | Automatic discharge after test ends (DCW) | |
| IR output voltage | 0.10 – 1.00 kV, ±2% | |
| IR resistance range | 0.1 MΩ – 10.0 GΩ | |
| IR accuracy | 0.10–999 MΩ: ±(10% rdg + 2 digits); 1.00–10.0 GΩ: ±(20% rdg + 2 digits) | |
| Arc-current detection range | 1 – 20 mA (separate sensing circuit) | |
| Voltage rise / fall time | 0.1 – 999.9 s | |
| Test time (AC/DC) | 0.2 – 999.9 s | |
| Waiting time (IR) | 0.2 – 999.9 s | |
| Time accuracy | ±(1% + 0.1 s) | |
| Memory | 20 test files, 20 test steps per file | |
| Dimensions (W × H × D) | 280 × 138 × 428 mm | |
| Weight | 18 kg | |
| Power supply | 100–121 V / 198–242 V AC (switchable), 47.5–63 Hz | |
Source: TH9320-S Series Datasheet, Rev. 2026-09. Specifications subject to change without notice.
Interfaces
- RS232standard
- USB Hoststandard
- USB Devicestandard
- Handlerstandard
- Signalstandard
- PLCstandard
- SCPI / MODBUSstandard
Accessories
Photos of fixtures and accessories: click an item code above, or browse all accessories.
Downloads
Interested in the TH9320-S Series?
Send us the model, quantity and your application. Our team in Ismaning will answer by email.